Rancang Bangun Alat Kalibrator Tensimeter NIBP Anak dan Dewasa

Authors

  • Muhammad Valen Brilian Sekolah Tinggi Ilmu Kesehatan Semarang
  • Mohamad Sofie Sekolah Tinggi Ilmu Kesehatan Semarang
  • Mohammad Rofi’i Sekolah Tinggi Ilmu Kesehatan Semarang

DOI:

https://doi.org/10.55606/jcsr-politama.v4i4.6524

Keywords:

Blood Pressure Calibration, Comparative Performance Test, NIBP Simulator, STM32F407VGT6, Tensimeter Calibrator

Abstract

Blood pressure is a vital parameter for diagnosing and monitoring cardiovascular conditions; therefore, the accuracy of digital blood pressure monitors (NIBP) must be maintained through periodic calibration using standard calibrators. However, commercial calibrators are relatively expensive and difficult to access, particularly for educational laboratories and resource-limited healthcare facilities. This study aimed to design, construct, and test the performance of an NIBP calibrator for pediatric and adult monitors—based on the STM32F407VGT6 microcontroller—capable of generating controlled blood pressure and heart rate simulations. The research methodology involved hardware and software design, circuit voltage measurement point testing, and a comparative performance test against both the developed NIBP simulator and a commercial NIBP simulator (used as a reference). Voltage test results aligned with component datasheet specifications. Performance testing involved five repetitions for systolic and diastolic parameters, comparing the developed calibrator with the commercial one. Results showed that all systolic and diastolic measurements fell within tolerance limits relative to standard settings, with a maximum average deviation of 2.2 mmHg. In the comparative test, all measurement points met the ±5 mmHg tolerance; the largest deviation was 5 mmHg (diastolic parameter at the highest pressure point of 100 mmHg), placing it exactly at the tolerance limit.

Downloads

Download data is not yet available.

References

(AAMI), A. f. (2022). ANSI/AAMI/ISO 81060-2:2019/(R)2022 Non-invasive Sphygmomanometers. Arlington: AAMI.

Indonesia, K. K. (2023). Pedoman Pengujian dan Kalibrasi Alat Kesehatan. Jakarta: Kemenkes RI.

Mahendra, A. R. (2024). Digital Pressure sebagai Alat Kalibrasi Tensimeter Digital, Tesis, Program Studi Teknik Elektromedik. Universitas Widya Husada Semarang.

Mahendra, R. (2024). “Rancang Bangun Alat Kalibrasi Tensimeter Digital Berbasis Arduino Uno,”. Jurnal Elektromedik Indonesia, vol. 6, no. 2, pp. 67–74.

Organization, W. H. (2022). Medical Device Technical Series: Blood Pressure Measuring Devices. WHO Press. Geneva: World Health Organization.

Pereira, J. M. (2025). “A Flexible, Low-Cost and Algorithm-Independent Calibrator for Automated Blood Pressure Measuring Devices,”. Applied Sciences, vol. 15, no. 6, pp. 1–16.

Pereira, L. S. (2025). “Low-Cost and Flexible Pressure Calibrator for Non-Invasive Blood Pressure Devices,”. Biomedical Engineering Online, vol. 24, no. 3, pp. 1–15.

Rahman, F. N. (2022). “Development of Digital Pressure Meter with Data Logging Feature for Medical Device Calibration,”. Journal of Instrumentation and Measurement, vol. 9, no. 1, pp. 33–40.

Siemasz, A. K. (2023). “Automatic Calibration of a Device for Blood Pressure Waveform Measurement,”. Sensors, vol. 23, no. 18, pp. 1–14.

Siemasz, A. K. (2023). “Automatic Calibration System for Non-Invasive Blood Pressure Measurement Devices,”. Measurement, vol. 210, pp. 112–120.

Standardization, I. O. (2022). ISO 81060-2: Non-Invasive Sphygmomanometers – Clinical Investigation of Automated Measurement Type. Geneva: International Organization for Standardization.

Standardization, I. O. (2023). Non-invasive sphygmomanometers – Requirements and test methods. ISO 81060-1, ISO 81060-1.

Utomo, A. H. (2025). “Design of Low-Cost Digital Pressure Meter for Digital Sphygmomanometer Calibration,” . Journal of Medical Instrumentation Technology, vol. 7, no. 1, pp. 45–52.

Utomo, H. d. (2022). “Digital Pressure Meter Design to Calibrate a Sphygmomanometer Unit Equipped with Data Calibration Recording to External Memory,”. Jurnal Teknokes, vol. 14, no. 2, pp. 85–92.

Utomo, H. H. ( (2025).). Design a low-cost digital pressure meter equipped with temperature and humidity parameters. Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics, 7(1), 1–8.

WHO. (2022). Medical Equipment Maintenance Programme Overview. WHO Press. Geneva: WHO.

Wibowo, A. P. (2025). Accuracy analysis on dual pressure (positive and negative) calibrator design to investigate the sensor response. journal of Electronics, Electromedical Engineering, and Medical Informatics, 7(1), 9–17.

Wibowo, R. P. (2025). “Accuracy Analysis of Dual-Pressure Calibrator for Blood Pressure Measurement Devices,”. International Journal of Biomedical Engineering, vol. 14, no. 2, pp. 101–109.

Downloads

Published

2026-07-28

How to Cite

Muhammad Valen Brilian, Mohamad Sofie, & Mohammad Rofi’i. (2026). Rancang Bangun Alat Kalibrator Tensimeter NIBP Anak dan Dewasa. Journal of Creative Student Research, 4(4), 214–224. https://doi.org/10.55606/jcsr-politama.v4i4.6524

Similar Articles

<< < 2 3 4 5 6 7 8 9 10 11 > >> 

You may also start an advanced similarity search for this article.